5.3 Replacing the 6V 30W Halogen Light
ALWAYS DISCONNECT YOUR EQUIPMENT BEFORE DOING ANY REPAIR.
Do not touch the glass envelope of the lamp during installation. Keep the protective
sleeve or bag of the lamp during installation and remove it right after installation.
LAMP AND LAMP HOUSING MAY BE HOT TO THE TOUCH.
• Switch off the microscope.
• Disconnect the power cord.
• Wait until the housing and
bulb have cooled sufficiently.
• Simply turn the illuminator
cover counterclockwise and
lift to remove.
• Remove the defective lamp.
• Place a new lamp into the
socket while avoiding
touching the glass. Notice the
bulb is wrapped in plastic to
avoid touching the glass
envelope.
• Be sure the bulb is pushed in
as far as possible thereby
aligning the filament in the illuminator.
• Reinstall the lamp housing making sure the notches on the housing and the cover
are lining up.
• Reconnect the power cord and switch on the microscope to verify proper
illuminator operation.